Retrieval of the Amplatzer duct occluder II device embolizing to the pulmonary artery by Goose Neck snare Pulmoner artere embolize olan Amplatzer duktus kapama II cihazının "Goose Neck" kement ile çıkarılması
Gulhane Medical Journal, cilt.52, sa.4, ss.276-279, 2010 (Scopus, TRDizin)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 52 Sayı: 4
- Basım Tarihi: 2010
- Dergi Adı: Gulhane Medical Journal
- Derginin Tarandığı İndeksler: Scopus, TR DİZİN (ULAKBİM)
- Sayfa Sayıları: ss.276-279
- Anahtar Kelimeler: "Goose neck" snare, Amplatzer duct occluder, Patent ductus arteriosus, Percutaneous transcatheter closure
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Percutaneous transcatheter closure of patent ductus arteriosus is a well-established procedure. Although it has been known as a safe treatment strategy, some complications have been reported. Embolization of the device, especially to the pulmonary artery, is one of these complications. In this report, we present a patient in whom an Amplatzer® duct occluder (ADO) II device embolized to the pulmonary artery and it was retrieved by using "Goose Neck" snare. We would like to notice the great importance of accurate preprocedural measurement of ductus arteriosus diameter during percutaneous transcatheter closure which seemsto be a simple procedure. © Gülhane Askeri Ti{dotless}p Akademisi 2010.