REDUCING THE NEED FOR REPEATING URINE DRUG TESTING WITH THE GRAY ZONE DETERMINED BY THE MEASUREMENT UNCERTAINTY SMANJENJE POTREBE ZA PONAVLJANJEM TESTIRANJA NA NEDOZVOLJENE SUPSTANCE U URINU U OKVIRU SIVE ZONE UZ POMO] MERNE NESIGURNOSTI


Creative Commons License

HUYSAL K., ÜSTÜNDAĞ Y., Çağlak H. A., Yeşil M. R.

Journal of Medical Biochemistry, cilt.42, sa.4, ss.616-620, 2023 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 42 Sayı: 4
  • Basım Tarihi: 2023
  • Doi Numarası: 10.5937/jomb0-41777
  • Dergi Adı: Journal of Medical Biochemistry
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, CAB Abstracts, Central & Eastern European Academic Source (CEEAS), Veterinary Science Database, Directory of Open Access Journals
  • Sayfa Sayıları: ss.616-620
  • Anahtar Kelimeler: cut-off, grey zone, measurement uncertainty, amphetamine, cannabinoid, cocaine
  • Açık Arşiv Koleksiyonu: AVESİS Açık Erişim Koleksiyonu
  • Sağlık Bilimleri Üniversitesi Adresli: Evet

Özet

Background: On an initial urine screening test for illegal substances, if the concentration of a substance is at or above the determined legislative threshold, it is reported as positive. Repeating testing with the same sample to verify it before reporting is a common practice in clinical laboratories. This study aimed to determine whether measurement uncertainty (MU) results can be used to detect a grey zone to reduce repeat testing. Methods: A retrospective study was conducted using data from the laboratory information system between January 1, 2020, and July 1, 2022. Samples studied twice within one hour before reporting for the same urine sample were analyzed. The MU values for urinary amphetamine, cannabinoid, cocaine, and opioid parameters were calculated using ADVIA Chemistry reagents on a Siemens ADVIA 1800 chemical analyzer. The grey zone was defined as the cut-off value ± MU. Results: Urine samples, 319 amphetamine, 198 cannabinoid, 112 cocaine, and 125 opiate tests were repeated from 31.839 patients (16–65 years). Ten amphetamine, 12 cannabinoid and 2 cocaine test results changed from positive to negative or negative to positive on retesting, all within the grey zone level defined by the MU value. Mean biases between the repeated test results were -7.64 (95% CI: -13.71 to -1.57) ng/mL for the amphetamine and 1.16 (95% CI: 0.31 to 2.01) ng/mL for the cannabinoid results. Conclusion: Repeating only values within the grey area for each illicit substance may be useful to reduce unnecessary duplicate measurements.