Scanning electron microscopy of removed lenses CIKARILAN INTRAOKULER LENSLERIN SCANNING ELEKTRON MIKROSKOPISI
Istanbul Tip Fakultesi Mecmuasi, cilt.59, sa.4, ss.86-90, 1996 (Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 59 Sayı: 4
- Basım Tarihi: 1996
- Dergi Adı: Istanbul Tip Fakultesi Mecmuasi
- Derginin Tarandığı İndeksler: Scopus
- Sayfa Sayıları: ss.86-90
- Anahtar Kelimeler: Intra ocular lenses, Scanning electron microscopy
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Surface analysis with scanning electron microscopy was performed on eleven removed from patients intraocular lenses. Two of them were anterior chamber and 9 of them were posterior chamber lenses. The surfaces of optics and haptics, optic-haptic junctions and positioning holes were examined. Manufacture defects were found in 8 lenses which the most frequently in optic-haptic junction and positioning holes and the less frequently in optic edge and haptics. Biodegration was seen in two polypropylene haptics and crater pattern was noted in one lens following Nd:YAG laser capsulotomy. Our findings were computed with causes of IOL removal in related literature. In conclusion the removal of IOLs will decrease with the use of monoblock IOLs without positioning holes manufactured under optimal technical conditions using a highly resistant material.