Subxiphoid approach to epicardial implantation of implantable cardioverter defibrillators in children


HAYDİN S., Saygi M., Ergul Y., Ozyilmaz I., ÖZTÜRK E., Akdeniz C., ...Daha Fazla

PACE - Pacing and Clinical Electrophysiology, cilt.36, sa.8, ss.926-930, 2013 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 36 Sayı: 8
  • Basım Tarihi: 2013
  • Doi Numarası: 10.1111/pace.12158
  • Dergi Adı: PACE - Pacing and Clinical Electrophysiology
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.926-930
  • Anahtar Kelimeler: implantable cardioverter defibrillator, pediatrics, surgery, subxiphoid
  • Sağlık Bilimleri Üniversitesi Adresli: Evet

Özet

Background Epicardial implantation of implantable cardioverter defibrillators (ICDs) is considered in the presence of intracardiac shunt, venous access issue, or small body size. We report our experience with epicardial ICD coil implantation using a minimally invasive method. Methods Nine patients who underwent epicardial ICD implantation were included. The median age was 7.4 years (3.9-9.6 years) and the median weight was 15 kg (12-24 kg). Diagnosis at the time of implantation included long QT syndrome (n = 5), catecholaminergic polymorphic ventricular tachycardia (n = 2), hypertrophic cardiomyopathy (n = 1), and fast monomorphic ventricular tachycardia (n = 1). The minimally invasive method involved a subxiphoid incision to place the epicardial pacing leads. The ICD coil was placed in the transverse sinus in four patients using an access path posterior to the heart. The second approach involved a path anterior to the heart to reach the epicardial location posterior to the left atrial appendage in the five remaining patients. No fluoroscopy was used in either approach. The median defibrillation threshold (DFT) at implantation was 7.5 J. Results Lower DFTs were observed in the anterior approach (10 J vs 5 J). Appropriate ICD shocks were observed in three patients during the median 18-month follow-up period (0.3-28 months). No inappropriate shocks were noted. One patient developed pericardial tamponade 39 days after the procedure and was surgically drained. Conclusions Minimally invasive epicardial ICD coil implantation in children with low DFT values is possible. The anterior implantation approach appears superior to the posterior approach. © 2013 Wiley Periodicals, Inc.