Percutaneous retrieval of an interatrial septal occluder device embolized into the aortic arch
Turk Kardiyoloji Dernegi Arsivi, cilt.38, sa.7, ss.502-504, 2010 (Scopus, TRDizin)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 38 Sayı: 7
- Basım Tarihi: 2010
- Dergi Adı: Turk Kardiyoloji Dernegi Arsivi
- Derginin Tarandığı İndeksler: Scopus, TR DİZİN (ULAKBİM)
- Sayfa Sayıları: ss.502-504
- Anahtar Kelimeler: Device removal/instrumentation, echocardiography, heart catheterization/complications, heart septal defects, atrial
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Percutaneous closure of secundum atrial septal defects (ASD) may be complicated by immediate embolization. We report on a 35-year-old woman who underwent percutaneous device closure for a secundum ASD. The diameter of the defect was measured as 4 mm by two-dimensional transesophageal echocardiography and a 7-mm Figulla ASD occluder device was implanted without prior balloon sizing of the defect. Immediate embolization was noted into the aortic arch. Attempts to pull the devices into the sheath with a loop snare failed even after replacing the delivery sheath with a bigger one. Finally, a bioptome was used to grab and place the screw mechanisms in the sheath and percutaneous retrieval of the embolized ASD occluder was achieved. Balloon sizing was performed after removal, yielding a stretched diameter of 12 mm and a 15-mm device was deployed with success.