Investigation of a Novel Method for High-Level Disinfection of the Tonometer Tip
Turkish Journal of Clinical and Experimental Ophthalmology, cilt.21, sa.1, ss.25-31, 2026 (Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 21 Sayı: 1
- Basım Tarihi: 2026
- Doi Numarası: 10.37844/tj-ceo.2026.21.4
- Dergi Adı: Turkish Journal of Clinical and Experimental Ophthalmology
- Derginin Tarandığı İndeksler: Scopus
- Sayfa Sayıları: ss.25-31
- Anahtar Kelimeler: disinfection, Goldmann tonometer, ozone
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Background/Aims: Disinfecting the tip of the Goldmann applanation tonometer with 3% hydrogen peroxide or 10% sodium hypochlorite for 10 minutes, followed by rinsing for 5 minutes, is not practical in busy clinics. This study aimed to explore a new, more practical method for disinfection. Methods: Ozone (O₃) gas (200 mg/h) and 20-watt, 411 nm blue light were used in the study. A sealed, light-and air-tight box containing an ozone generator and blue light source was used. Petri dishes (r = 9 cm) were inoculated with E. coli ATCC 25922 at concentrations of 10⁷ CFU/mL and 10⁸ CFU/mL by diluting a 0.5 McFarland suspension 1:10, and applying it evenly with a swab. To observe physical effects, 1 cm² polymethyl methacrylate (PMMA) plates were also placed inside the box. Experiments were conducted using two bacterial densities (10⁷ and 10⁸ CFU/mL), with three ozone doses (1.335, 2.670, and 5.340 mg/L) and three blue light exposure times (8, 16, and 32 minutes). Each test was repeated twice. Results: With 10⁷ CFU/mL bacteria, 1.335 mg/L ozone and 8 minutes of blue light reduced colonies by 98.2%; after 32 minutes, no colonies remained. At 10⁸ CFU/mL, 5.340 mg/L ozone and 32 minutes of light reduced colonies by 98.5%. Conclusions: The combination of ozone and 411 nm blue light achieved dose-dependent disinfection even at high bacterial concentrations (10⁸ CFU/mL). This method appears to be effective under experimental conditions.