Investigation of a Novel Method for High-Level Disinfection of the Tonometer Tip


ŞANAL DOĞAN A., Kocamis S. İ., Kosak G., Coban A., Morkoc M., Erdem G., ...Daha Fazla

Turkish Journal of Clinical and Experimental Ophthalmology, cilt.21, sa.1, ss.25-31, 2026 (Scopus)

Özet

Background/Aims: Disinfecting the tip of the Goldmann applanation tonometer with 3% hydrogen peroxide or 10% sodium hypochlorite for 10 minutes, followed by rinsing for 5 minutes, is not practical in busy clinics. This study aimed to explore a new, more practical method for disinfection. Methods: Ozone (O₃) gas (200 mg/h) and 20-watt, 411 nm blue light were used in the study. A sealed, light-and air-tight box containing an ozone generator and blue light source was used. Petri dishes (r = 9 cm) were inoculated with E. coli ATCC 25922 at concentrations of 10⁷ CFU/mL and 10⁸ CFU/mL by diluting a 0.5 McFarland suspension 1:10, and applying it evenly with a swab. To observe physical effects, 1 cm² polymethyl methacrylate (PMMA) plates were also placed inside the box. Experiments were conducted using two bacterial densities (10⁷ and 10⁸ CFU/mL), with three ozone doses (1.335, 2.670, and 5.340 mg/L) and three blue light exposure times (8, 16, and 32 minutes). Each test was repeated twice. Results: With 10⁷ CFU/mL bacteria, 1.335 mg/L ozone and 8 minutes of blue light reduced colonies by 98.2%; after 32 minutes, no colonies remained. At 10⁸ CFU/mL, 5.340 mg/L ozone and 32 minutes of light reduced colonies by 98.5%. Conclusions: The combination of ozone and 411 nm blue light achieved dose-dependent disinfection even at high bacterial concentrations (10⁸ CFU/mL). This method appears to be effective under experimental conditions.