Comparison of Intraocular Pressure and Pachymeter Measurements Obtained by Non Contact Tonometer-pachymeter with Goldmann Applanation Tonometer and Corneal Topography
Turkish Journal of Clinical and Experimental Ophthalmology, cilt.16, sa.1, ss.12-17, 2021 (Scopus, TRDizin)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 16 Sayı: 1
- Basım Tarihi: 2021
- Doi Numarası: 10.37844/glauc.cat.2021.16.3
- Dergi Adı: Turkish Journal of Clinical and Experimental Ophthalmology
- Derginin Tarandığı İndeksler: Scopus, TR DİZİN (ULAKBİM)
- Sayfa Sayıları: ss.12-17
- Anahtar Kelimeler: Central corneal thickness, Intraocular pressure, Non-contact tono-pachymeter
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Purpose: To compare the measurements of intraocular pressure (IOP) and central corneal thickness (CCT) by tono-pachymeter (CT-1P) with IOP measurements by Goldmann applanation tonometer (GAT), and CCT measurements by Pentacam. Materials and Methods: Two hundred eighteen eyes of 109 patients were included in the study. Intraocular pressure measurements were obtained by the CT-1P and GAT, whereas CCT measurements were obtained by CT-1P and Pentacam. Blant-Altman plots for agreement, and Pearson correlation and interclass correlation (ICC) for reliability between the measurements were analyzed. Results: The mean IOPCT-1P, IOPCT-1P adj., IOPGAT were measured as 18.2±4.32 mmHg, 17.75±4.05 mmHg and 16.9±4.16 mmHg, respectively. The mean CCT was measured as 554.8±37.3 μm and 543.6±35.9 μm with CT-1P and Pentacam, respectively. Mean differences between devices were significantly different in terms of IOP and CCT measurement (p<0.001, for all). Interclass correlation coefficient confirmed the excellent reliability between devices in terms of measurement of IOP and CCT (p<0.001, for all). The 95% limits of agreement (LoA) for IOP were-4.24 to 3.28 mmHg between IOPCT-1P adj. and IOPCT-1P,-3.49 to 5.14 mmHg between IOPCT-1P adj. and IOPGAT and-2.7 to 5.3 mmHg between IOPCT-1P and IOPGAT. Also 95% LoA for CCT were-12.33 to 32.16 μm between CCTCT-1P and CCTPentacam. Conclusion: Although CT-1P overestimated both IOP and CCT measurements, the device had an excellent ICC with Pentacam and GAT in terms of CCT and IOP measurements. The ability of tono-pachymeters measuring both CCT and IOP with a single device provide an important convenience in busy clinics.