Lesion detectability and clinical effectiveness of dual-head coincidence gamma camera imaging in comparison with dedicated PET systems in tumour patients


Narin Y., URHAN M., Canpolat N., Vardereli E., Bayhan H.

Journal of International Medical Research, cilt.35, sa.4, ss.467-473, 2007 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 35 Sayı: 4
  • Basım Tarihi: 2007
  • Doi Numarası: 10.1177/147323000703500405
  • Dergi Adı: Journal of International Medical Research
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.467-473
  • Anahtar Kelimeler: cancer, dual-head coincidence gamma camera, didicated PET, diagnostic imaging
  • Sağlık Bilimleri Üniversitesi Adresli: Evet

Özet

The lesion detection capability and clinical effectiveness of dual-head coincidence gamma camera imaging (c-PET) were compared with those of dedicated positron emission tomography (d-PET) in 37 cancer patients who underwent whole-body c-PET and d-PET imaging after administration of 370 - 540 MBq 18F-fluorodeoxyglucose. Eighty-nine lesions were detected on c-PET whereas 133 lesions were seen with d-PET imaging. The relative sensitivity of c-PET compared with d-PET was 62% and 73% for lesions < 15 and ≥ 15 mm, respectively, and the relative concordance rate was 84% when the patients were restaged. Since the lesion detection rate of c-PET imaging was lower than that of d-PET, the detection of small lesions, therefore, requires care. The clinical effectiveness of c-PET, however, was similar to that of d-PET and, therefore, it is concluded that c-PET can be used as an alternative to d-PET, particularly considering the high cost and limited availability of d-PET cameras. Copyright © 2007 Field House Publishing LLP.