Comparison of the INR Values Measured by CoaguChek XS Coagulometer and Conventional Laboratory Methods in Patients on VKA Therapy
Clinical and Applied Thrombosis/Hemostasis, cilt.23, sa.2, ss.187-194, 2017 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 23 Sayı: 2
- Basım Tarihi: 2017
- Doi Numarası: 10.1177/1076029615595881
- Dergi Adı: Clinical and Applied Thrombosis/Hemostasis
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
- Sayfa Sayıları: ss.187-194
- Anahtar Kelimeler: international normalized ratio, oral anticoagulation, point-of-care systems, warfarin
- Sağlık Bilimleri Üniversitesi Adresli: Evet
Özet
Background: Warfarin, which is a widely used oral anticoagulant, has a narrow therapeutic window and requires regular international normalized ratio (INR) monitoring to maintain optimal anticoagulation. Recently, several portable coagulometers have been developed to measure INR levels. Objective: To compare the INR results obtained by a portable coagulometer (CoaguChek XS) and a standard laboratory method (STAGO STA-R). Methods: Overall, 433 consecutive patients (male: 191, median age: 61 [44-86] years) who were admitted to outpatient anticoagulation clinic were enrolled in this study. Each patient was tested for INR using portable CoaguChek XS and STAGO STA-R automatic laboratory coagulometer. Correlation between methods was assessed using the Pearson correlation test and Cohen κ test. Bland-Altman plot was used to identify mean difference and 95% limits of agreement. Results: The mean INR values for CoaguChek XS and STAGO STA-R were 2.54 ± 1.17 and 2.79 ± 1.39, respectively. There was a strong positive correlation between the 2 methods (r =.966; 95% confidence interval [CI]: 0.95-0.97, P <.001). The Bland-Altman analysis gave a mean difference of 0.26 ± 0.40 between the 2 methods, with a 95% limit of agreement of -0.54 to 1.05. In patients with INR values >5.0, there was only a moderate correlation (r =.676; 95% CI: 0.38-0.89, P =.002), and the mean difference of INR tended to increase as mean INR values increased. There was a high overall agreement between the 2 methods (κ =.751; 95% CI: 0.69-0.80; P <.001). Conclusion: There was good consistency between traditional laboratory testing and CoaguChek XS coagulometer, which provides rapid and reliable INR analysis.